Calibration of sound calibrators
RISE is calibrating all types of sound calibrators
Purpose/Benefit:The national laboratories are responsible for maintaining the national reference standards for each measurement quantity and spreading metrological competence and traceability to industry and society at large. The national reference standards are the highest level of traceability in Sweden.
A calibration performed at a national laboratory fulfils all typical traceability and accreditation requirements that commonly exist in industry and society at large.
Method (what/which methods are used to perform the service):The characteristics of sound calibrators are specified in the international standard IEC 60942. Requirements and tolerances for sound pressure level, frequency, distortion and stability are among other things stated in this standard. The standard gives three classes of accuracy, LS, 1 and 2. Class LS is intended only for laboratory use while class 1 and 2 are intended for field use. These classes match the corresponding class for sound level meters.
RISE calibrates all types of sound calibrators. Parameters which are calibrated are sound pressure level, frequency and distortion. If the instrument is equipped with an accessory barometer, we check this as well at current air pressure
The recommended calibration interval is 1 year for sound calibrators.
Delivery (what does the client get after performed service – e.g. a report, certificate etc.):The calibration results are reported in a digital calibration certificate in Swedish or English which states that the calibration was performed at a national laboratory.
Area:Calibration
Generic metrology and measurement technology Contact person (Enter one name per field. Activated personal contact pages will appear automatically):
Camilla Ekström, Ingenjör
Håkan Andersson, FoU ingenjör
If the sound calibrator is configured for an other microphone size than ½", please enclose the ½" adapter if available.
Link to order form: Website Calibration Standards:IEC 60942
Certification and marking: Not applicable Type of service: Calibration Instrument: Acoustic measurement equipment General area: Sound and vibrations Order information: Order can be sent to risekalibrering@ri.se, or through the case management in our customer website Calibration. Divison (OLD): Do not use - Division Built Environment Delivery level: National Metrology InstituteTotal defence and crisis preparedness
Health and life science Tjänstetyp tagg: Kalibrering