Calibration of sound calibrators

RISE is calibrating all types of sound calibrators


This service is also relevant in: Total defence and crisis preparedness, Health and life science

Service details

Service
Calibration of sound calibrators
Instrument
Acoustic measurement equipment
Metrology Area
Sound and vibrations
Delivery level
National Metrology Institute
Standards

IEC 60942

Price
Price on tender
Preparations

If the sound calibrator is configured for an other microphone size than ½", please enclose the ½" adapter if available.


Contact us

CAPTCHA
By submitting the form, RISE will process your personal data.

Camilla Ekström

Ingenjör
+46 10 516 59 43 Read more about Camilla

Contact Camilla

CAPTCHA

* Mandatory 

By submitting the form, RISE will process your personal data.

Håkan Andersson

FoU ingenjör
+46 10 516 54 23 Read more about Håkan

Contact Håkan

CAPTCHA

* Mandatory 

By submitting the form, RISE will process your personal data.

Purpose

The national laboratories are responsible for maintaining the national reference standards for each measurement quantity and spreading metrological competence and traceability to industry and society at large. The national reference standards are the highest level of traceability in Sweden.

A calibration performed at a national laboratory fulfils all typical traceability and accreditation requirements that commonly exist in industry and society at large. 

Method

The characteristics of sound calibrators are specified in the international standard IEC 60942. Requirements and tolerances for sound pressure level, frequency, distortion and stability are among other things stated in this standard. The standard gives three classes of accuracy, LS, 1 and 2. Class LS is intended only for laboratory use while class 1 and 2 are intended for field use. These classes match the corresponding class for sound level meters.

RISE calibrates all types of sound calibrators. Parameters which are calibrated are sound pressure level, frequency and distortion. If the instrument is equipped with an accessory barometer, we check this as well at current air pressure

The recommended calibration interval is 1 year for sound calibrators.

Deliveries

The calibration results are reported in a digital calibration certificate in Swedish or English which states that the calibration was performed at a national laboratory.