Calibration and verification of sound level meters
RISE is verifying most types of sound level meters and sound analysers.
Purpose/Benefit:The national laboratories are responsible for maintaining the national reference standards for each measurement quantity and spreading metrological competence and traceability to industry and society at large. The national reference standards are the highest level of traceability in Sweden.
A calibration performed at a national laboratory fulfils all typical traceability and accreditation requirements that commonly exist in industry and society at large.
Method (what/which methods are used to perform the service):The characteristics of sound level meters are specified in the international standard IEC 61672, or for older instruments IEC 60651 and IEC 60804. Measurement procedures and tolerances to verify that the characteristics of the sound level meter are correct are also stated in these standards. Examples of characteristics which are tested: frequency response of microphone and filter, the different time weightings (Fast, Slow etc.), level linearity, overrange indication etc.
Sound level meters are divided into different classes of accuracy, where class 1 is normal for precision measurements and class 2 for simpler measurements. Approximate accuracy for a sound level meter of class 1 is 1 dB and 2 dB for class 2. However, the accuracy is dependent on the character of the sound.
Sound level meters have often an octave band analysis feature. Characteristics for this is specified in a separate standard, IEC 61260.
Delivery (what does the client get after performed service – e.g. a report, certificate etc.):The calibration results are reported in a digital calibration certificate in Swedish or English which states that the calibration was performed at a national laboratory.
Area:Calibration
Generic metrology and measurement technology Contact person (Enter one name per field. Activated personal contact pages will appear automatically):
Camilla Ekström, Ingenjör
Håkan Andersson, FoU ingenjör
IEC 61672
IEC 61260
Certification and marking: Not applicable Type of service: Calibration Instrument: Acoustic measurement equipment General area: Sound and vibrations Order information: Order can be sent to risekalibrering@ri.se, or through the case management in our customer website Calibration. Divison (OLD): Do not use - Division Built Environment Delivery level: National Metrology InstituteRisk and security
Production and manufacturing
Health and life science Tjänstetyp tagg: Kalibrering