Jump directly to content

Testing of magnetic field meters

Testing of magnetic field meters/field probes in the frequency range 5 Hz-1 MHz with field strength up to 10 uT


This service is also relevant in: Sensors and sensor systems, Digital infrastructure

Service details

Service
Testing of magnetic field meters/field probes in the frequency range 5 Hz-1 MHz
Instrument
Field meters and probes
Metrology Area
Electricity
Price
Price on tender

Order

CAPTCHA
By submitting the form, RISE will process your personal data.

Mats Cedheim

Ingenjör
+46 10 516 60 86 Read more about Mats

Contact Mats

CAPTCHA

* Mandatory 

By submitting the form, RISE will process your personal data.

Fredrik Harrysson

Enhetschef
Read more about Fredrik

Contact Fredrik

CAPTCHA

* Mandatory 

By submitting the form, RISE will process your personal data.

Purpose

To provide testing of magnetic field meters/field probes in the frequency range 5 Hz-1 MHz for field strength up to 10 uT 

The correction factor is used to improve the accuracy of the tested equipment.

Method

We test the equipment according to our internal metod 5031. This is done by placing the magnetic field meter/field probe in a calcuable known field created with a Helmholtz coil. The quota between the calculated applied field and the readout from the equipment gives a correction factor unique´for the specific equipment at the given frequency. 

Deliveries

Report containing the correction factors at requested frequencies