Calibration of 633 nm HeNe-lasers at The National Laboratory
The National Laboratory for Length and Dimensional Metrology offers calibration of HeNe-Laser sources at 633 nm.
This service is also relevant in: Fiber optics and photonics, Metrology, Energy and electrification
Service details
- Service
- Calibration of 633 nm HeNe-lasers at The National Laboratory
- Instrument
- Lasers
- Metrology Area
- Length and geometry
- Delivery level
- National Metrology Institute
- Price
Price on request.
Purpose
The national laboratories are responsible for maintaining the national reference standards for each measurement quantity and spreading metrological competence and traceability to industry and society at large. The national reference standards are the highest level of traceability in Sweden.
A calibration performed at a national laboratory fulfils all typical traceability and accreditation requirements that commonly exist in industry and society at large.
Method
Calibration of the object’s laser frequency is performed via heterodyne measurement against one of Sweden’s primary standards for optical frequency. The standards are Iodine-stabilized He-Ne lasers with a laser frequency in agreement with BIPMs recommended values for the realization of the metre in the mise en pratique.
The results from the calibration are:
- Measured vacuum wavelength
- Relative deviation from the specified vacuum wavelength
- Difference between longest and shortest measured wavelength
- Diagram showing the relative deviation of the vacuum wavelength from specified vacuum wavelength
Additionally the optical power is measured.
The laser frequency calibration certificate is issued under the CIPM MRA.
Deliveries
The calibration results are reported in a digital calibration certificate in Swedish or English which states that the calibration was performed at a national laboratory.
When becoming a client you will also get access to our customer website for calibration, where you may also place orders.